  Control N (Lot No.: AE0909N) Control D (Lot No.: AE0909D)
UnitID Reagent Kit
(Code)
Target
(U/gHb)
Mean
(U/gHb)
n for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ  Target
(U/gHb)
Mean
(U/gHb)
n for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ
Lab003 1 14.414.125 0.53.59.2205.1   5.85.825 0.23.46.9205.9
Lab006 1 14.413.85 0.53.611.4204.4   5.85.65 0.11.87.020>6
Lab007 1 14.314.513 0.42.86.920>6   5.65.713 0.11.85.320>6
Lab009 1 14.414.15 0.21.44.920>6   5.85.75 0.11.85.220>6
Lab011 1 14.413.65 0.21.58.520>6   5.85.75 0.11.85.220>6
Lab012 1 14.414.26 0.64.29.8204.4   5.96.06 0.35.011.7203.7
Lab013 1 14.513.69 0.42.912.1204.8   5.75.79 0.11.83.520>6
Lab014 1 14.414.110 0.32.16.320>6   5.85.710 0.23.58.7205.2
Lab018 2 15.710.87 0.54.640.520-2.4   5.64.27 0.49.544.020-0.5
Lab029 1 14.414.54 0.10.72.120>6   5.85.84 0.11.73.420>6
誤差 (Bias)% = { [ │累計平均值 (Mean) - 目標值 (Target)│] / 目標值 (Target) } x 100%
總誤差 (TE, Total Error)% = 誤差 (Bias)% + 2 × 累計變異係數 (CV)
σ (Sigma) = (TEa% - 誤差 (Bias)%) / 累計變異係數 (CV)
[TOP]



累計統計表

累計期間:自
  Control N (Lot No.: AE0909N) Control D (Lot No.: AE0909D)
UnitID Reagent Kit
(Code)
Target
(U/gHb)
Mean
(U/gHb)
n for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ  Target
(U/gHb)
Mean
(U/gHb)
n for
Mean
SDCV
(%)
TE
(%)
TEa
(%)
σ
Lab003 1,6 14.414.4268 0.42.85.620>6   5.85.8268 0.23.46.9205.9
Lab004 1 14.814.510 0.32.16.220>6   5.75.710 0.11.83.520>6
Lab006 1 14.413.749 0.42.910.7205.2   5.85.649 0.23.610.6204.6
Lab007 1 14.313.9285 0.96.515.7202.6   5.65.4285 0.47.418.4202.2
Lab009 1 14.414.576 0.96.213.1203.1   5.85.976 0.46.815.3202.7
Lab010 1 15.315.519 0.53.27.8205.8   6.36.519 0.34.612.4203.7
Lab011 1 14.413.753 0.32.29.220>6   5.85.753 0.11.85.220>6
Lab012 1 14.414.393 0.42.86.320>6   5.95.993 0.23.46.8205.9
Lab013 1,3 14.513.844 0.42.910.6205.2   5.75.744 0.11.83.520>6
Lab014 1 14.414.587 0.53.47.6205.7   5.85.687 0.23.610.6204.6
Lab016 1,3 15.014.834 0.85.412.1203.5   5.35.434 0.35.613.0203.2
Lab018 2,5 15.711.822 2.319.563.820-0.2   5.64.522 0.817.855.2200.0
Lab029 1,4 14.414.432 0.21.42.820>6   5.85.732 0.11.85.220>6
Lab030 1 14.414.33 0.53.57.7205.5   5.85.43 0.35.618.0202.3
Lab032 2 11.512.46 0.64.817.5202.5   4.54.86 0.36.319.2202.1
Lab043 2,5 11.511.29 1.513.429.4201.3   4.54.29 0.511.930.5201.1
Lab044 1,3 14.415.258 0.95.917.4202.4   5.86.358 0.46.321.3201.8
誤差 (Bias)% = { [ │累計平均值 (Mean) - 目標值 (Target)│] / 目標值 (Target) } x 100%
總誤差 (TE, Total Error)% = 誤差 (Bias)% + 2 × 累計變異係數 (CV)
σ (Sigma) = (TEa% - 誤差 (Bias)%) / 累計變異係數 (CV)