Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | BA1116N | BA1116D |
Duration of the Analyzing | Month (2024/11) | CUM (2022/06/01~2024/11/30) | Month (2024/11) | CUM (2022/06/01~2024/11/30) |
Runs (N) | 4 | 53 | 4 | 53 |
Mean (U/gHb) | 14.4 | 14.6 | 4.6 | 5.2 |
SD | 0.4 | 0.4 | 0.2 | 0.5 |
CV (%) | 2.8 | 2.7 | 4.3 | 9.6 |
Target Value (U/gHb) | 14.4 | 14.4 | 5.7 | 5.7 |
Total Error (%) | 5.6 | 6.9 | 28.0 | 28.0 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | >6 | >6 | 0.2 | 1.2 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: BA1116N ; Duration : 2022-06-01 to 2024-11-30 ; Target : 14.4 ; SD : 0.50 ( ●) | |
Lot No.: BA1116D ; Duration : 2022-06-01 to 2024-11-30 ; Target : 5.7 ; SD : 0.20 ( □) | |
|
Lot No.: BA1116N ; Duration : 2022-06-01 to 2024-11-30 ; Target : 14.4 ; SD : 0.50 | |
|
Lot No.: BA1116D ; Duration : 2022-06-01 to 2024-11-30 ; Target : 5.7 ; SD : 0.20 | |
|
Contact us
Address:
P.O. Box 624 Taipei Xinwei, Taiwan, 106926 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 09 Apr, 2020