Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | BA1116N | BA1116D |
Duration of the Analyzing | Month (2025/02) | CUM (2022/06/01~2025/02/28) | Month (2025/02) | CUM (2022/06/01~2025/02/28) |
Runs (N) | 4 | 26 | 4 | 26 |
Mean (U/gHb) | 14.3 | 13.8 | 4.4 | 4.3 |
SD | 0.1 | 0.4 | 0.1 | 0.2 |
CV (%) | 0.7 | 2.9 | 2.3 | 4.7 |
Target Value (U/gHb) | 14.9 | 14.9 | 5.1 | 5.1 |
Total Error (%) | 5.4 | 13.2 | 18.3 | 25.0 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | >6 | 4.4 | 2.7 | 0.9 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: BA1116N ; Duration : 2022-06-01 to 2025-02-28 ; Target : 14.9 ; SD : 2.30 ( ●) | |
Lot No.: BA1116D ; Duration : 2022-06-01 to 2025-02-28 ; Target : 5.1 ; SD : 2.50 ( □) | |
 |
Lot No.: BA1116N ; Duration : 2022-06-01 to 2025-02-28 ; Target : 14.9 ; SD : 2.30 | |
 |
Lot No.: BA1116D ; Duration : 2022-06-01 to 2025-02-28 ; Target : 5.1 ; SD : 2.50 | |
 |