Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | BA1116N | BA1116D |
Duration of the Analyzing | Month (2024/11) | CUM (2022/06/01~2024/11/30) | Month (2024/11) | CUM (2022/06/01~2024/11/30) |
Runs (N) | 3 | 52 | 3 | 52 |
Mean (U/gHb) | 15.5 | 14.9 | 4.7 | 4.8 |
SD | 0.2 | 0.4 | 0.1 | 0.6 |
CV (%) | 1.3 | 2.7 | 2.1 | 12.5 |
Target Value (U/gHb) | 14.9 | 14.9 | 4.8 | 4.8 |
Total Error (%) | 6.6 | 5.4 | 6.3 | 25.0 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | >6 | >6 | >6 | 1.6 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: BA1116N ; Duration : 2022-06-01 to 2024-11-30 ; Target : 14.9 ; SD : 1.28 ( ●) | |
Lot No.: BA1116D ; Duration : 2022-06-01 to 2024-11-30 ; Target : 4.8 ; SD : 0.42 ( □) | |
|
Lot No.: BA1116N ; Duration : 2022-06-01 to 2024-11-30 ; Target : 14.9 ; SD : 1.28 | |
|
Lot No.: BA1116D ; Duration : 2022-06-01 to 2024-11-30 ; Target : 4.8 ; SD : 0.42 | |
|