Internal Quality Control Program (IQC) for G6PD Quantitative Test
G6PD IQC MIS System [login]
Supported by Preventive Medicine Foundation
| Control N | Control D |
QC Control Lot No. | BA1116N | BA1116D |
Duration of the Analyzing | Month (2025/02) | CUM (2022/06/01~2025/02/28) | Month (2025/02) | CUM (2022/06/01~2025/02/28) |
Runs (N) | 1 | 56 | 1 | 56 |
Mean (U/gHb) | 14.9 | 15.0 | 5.2 | 4.8 |
SD | - | 0.4 | - | 0.5 |
CV (%) | - | 2.7 | - | 10.4 |
Target Value (U/gHb) | 14.9 | 14.9 | 4.8 | 4.8 |
Total Error (%) | - | 6.0 | - | 20.8 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | - | >6 | - | 1.9 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Control N and Control D SDI QC Chart |
Lot No.: BA1116N ; Duration : 2022-06-01 to 2025-02-28 ; Target : 14.9 ; SD : 1.28 ( ●) | |
Lot No.: BA1116D ; Duration : 2022-06-01 to 2025-02-28 ; Target : 4.8 ; SD : 0.42 ( □) | |
 |
Lot No.: BA1116N ; Duration : 2022-06-01 to 2025-02-28 ; Target : 14.9 ; SD : 1.28 | |
 |
Lot No.: BA1116D ; Duration : 2022-06-01 to 2025-02-28 ; Target : 4.8 ; SD : 0.42 | |
 |
Contact us
Address:
P.O. Box 624 Taipei Xinwei, Taiwan, 106926 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 09 Apr, 2020